Data File : C:\Users\P...ion\1\Data\Multi Sample 2023-02-03 11-29-48\F-002-102-3 standards.D Acq. Method: Terps and Cannabs.M The Acq. Method's Instrument Parameters for the Run were : ===================================================================== Agilent 8890 ===================================================================== GC GC Summary Run Time 20 min Post Run Time 0 min Oven Equilibration Time 1 min Max Temperature 350 °C Maximum Temperature Override Disabled Slow Fan Disabled Temperature Setpoint On (Initial) 40 °C Hold Time 3.2 min Post Run 50 °C Program #1 Rate 25 °C/min #1 Value 250 °C #1 Hold Time 8.4 min ALS Front Injector Syringe Size 10 μL Injection Volume 1 μL Solvent A Washes (PreInj) 3 Solvent A Washes (PostInj) 3 Solvent A Volume 8 μL Solvent B Washes (PreInj) 0 Solvent B Washes (PostInj) 0 Solvent B Volume 8 μL Sample Washes 2 Sample Wash Volume 2 μL Sample Pumps 6 Dwell Time (PreInj) 0 min Dwell Time (PostInj) 0 min Solvent Wash Draw Speed 300 μL/min Solvent Wash Dispense Speed 3000 μL/min Sample Wash Draw Speed 300 μL/min Sample Wash Dispense Speed 3000 μL/min Injection Dispense Speed 6000 μL/min Viscosity Delay 2 sec Sample Depth Disabled Injection Type Standard L1 Airgap 0.2 μL Solvent Wash Mode A, B Sample Overlap Mode Sample overlap is not enabled Front SS Inlet H2 Mode Split Heater On 250 °C Pressure On 4.4452 psi Total Flow On 38.7 mL/min Septum Purge Flow On 3 mL/min Pre-Run Flow Test Off Gas Saver On 20 After 2 min mL/min Split Ratio 50 :1 Split Flow 35 mL/min Liner A Liner has not been selected. Column Column Outlet Pressure 0 psi Column #1 Column Information Agilent 19091S-433UI: T330351H HP-5MS UI Temperature Range -60 °C—325 °C (350 °C) Dimensions 30 m x 250 μm x 0.25 μm Column lock Locked In Front SS Inlet H2 Out Front Detector FID (Initial) 40 °C Pressure 4.4452 psi Flow 0.7 mL/min Average Velocity 21.646 cm/sec Holdup Time 2.3099 min Control Mode Constant Flow Flow Setpoint On (Initial) 0.7 mL/min Post Run 1 mL/min Front Detector FID Makeup N2 Heater On 300 °C H2 Flow (Combined) On 30 mL/min Air Flow On 400 mL/min Makeup Flow On 20 mL/min Carrier Gas Flow Correction Column + Fuel = Constant Flame On Blank Evaluation Setpoints Perform Blank Evaluation Test Off Initial Baseline Minimum 2 pA Initial Baseline Maximum 20 pA Initial Baseline Noise 0.3 pA Final Baseline Minimum 2 pA Final Baseline Maximum 40 pA Final Baseline Noise 0.6 pA Total Peak Area 100 pA*sec Maximum Peak Height 3 pA Time Window Start 0 min Time Window End 8.2 min Detector Evaluation Perform Detector Evaluation Test Off Signal Selected Signal #1: Front Signal (FID) {5 Hz} Checkout Sample FID Signals Signal #1: Front Signal Description Front Signal Details Front Signal (FID) Save On Data Rate 5 Hz Signal #2: Description None Signal #3: Description None Signal #4: Description None Signal #5: Description None Signal #6: Description None Signal #7: Description None Signal #8: Description None ===================================================================== Column(s) ===================================================================== Column Description : HP-5MS UI Inventory# : T330351H Model# : 19091S-433UI Manufacturer : Agilent Diameter : 250.0 µm Length : 30.0 m Film thickness : 0.25 µm Void time : 2.310 min Maximum Temperature: 325.0 °C Comment :